Objective: The aim of this study was to compare the consistency of objective optometry results between VX120 multi-diagnostic unit and TopconKR800 automatic optometer. Methods: A total of 24 patients (24 right eyes) admitted to our hospital from November 2019 to December 2019 for corneal refractive surgery were collected. Optometry was performed on 24 patients with TopconKR800 automatic optometer and VX120 multi-diagnostic unit, and got the results following: spherical mirror photometry (S), cylindrical lens luminosity (C), mean spherical equivalent (M), horizontal and vertical cylindrical lens luminosity ( J0) and 45° cylindrical lens luminosity ( J45). Compared T-test analysis was used to compare the parameters of the two devices, and the Bland-Altman 95% limits of agreement were performed. Results: The S measured by the VX120 multi-diagnostic unit with bright mode was (-3.90±3.06) D; M was (-4.41±3.24) D; J0 was (0.34±0.58) D;J45 was (-0.04±0.26) D. The S measured by TopconKR800 automatic optometer was (-3.77±2.88) D; M was (-4.15±3.07) D; J0 was (0.25±0.47) D; J45 was (-0.01±0.18) D. There was no statistically significant difference in S, M, and J45 measured by VX120 multi-diagnostic unit and TopconKR800 automatic optometer in bright mode, while there was statistically significant difference in J0 between VX120 multi-diagnostic unit and TopconKR800 automatic optometer (t=-2.24, P=0.04). Bland-Altman 95% consistency analysis showed that the data points of the two instruments were mostly distributed within the 95% confidence interval, indicating that the parameters measured by these two devices were in accord with each other. Conclusion: The objective optometry results of VX120 multi-diagnostic unit and TopconKR800 automatic optometer were consistent and can be substituted for each other in the refraction examination.
Lai YH, Tseng HY, Hsu HT, et al. Uncorrected visual acuity and
noncycloplegic autorefraction predict significant refractive errors in
Taiwanese preschool children[J]. Ophthalmology, 2013, 120(2): 271-276.
Harvey EM, Miller JM, Schwiegerling J. Utility of an open field ShackHartmann aberrometer for measurement of refractive error in infants
and young children[ J]. J AAPOS, 2013, 17(5): 494-500.
He M, Xiang F, Zeng Y, et al. Effect of time spent outdoors at school on
the development of myopia among children in China: a randomized
clinical trial[ J]. JAMA, 2015, 314(11): 1142-1148.
Dirani M, Chan YH, Gazzard G, et al. Prevalence of refractive error
in Singaporean Chinese children: the strabismus, amblyopia, and
refractive error in young Singaporean Children (STARS) study[ J].Invest Ophthalmol Vis Sci, 2010, 51(3): 1348-1355.
Kuo YC, Wang JH, Chiu CJ. Comparison of open-field autorefraction,
closed-field autorefraction, and retinoscopy for refractive measurements
of children and adolescents in Taiwan[ J]. J Formos Med Assoc, 2020,
119(8): 1251-1258.
Gordon-Shaag A, Pi?ero DP, Kahloun C, et al. Validation of refraction
and anterior segment parameters by a new multi-diagnostic platform
(VX120)[ J]. J Optom, 2018, 11(4): 242-251.
Pi?ero DP, López-Navarro A, Cabezos I, et al. Intrasession repeatability
of refractive and ocular aberrometric measurements obtained using a
multidiagnostic device in healthy eyes[ J]. Clin Optom (Auckl), 2017,
9: 91-96.
Sachdev N, Cairns G, McGhee CN. A comparison of autorefractor
performance[ J]. Optom Vis Sci, 2005, 82(1): 9; author reply 9-10.
卢纯洁, 宋迪, 陈仪泽, 等. 三种自动验光仪检测儿童青少年屈光
不正的准确性和一致性比较[ J]. 中华眼视光学与视觉科学杂
志, 2019, 21(10): 721-726.
LU CJ, SONG D, CHEN YZ, et al. Comparison of refractive
measures of three autorefractors in school-aged children and
adolescents[ J]. Chinese Journal of Optometry Ophthalmology and
Visual Science, 2019, 21(10): 721-726.
Sun YY, Wei SF, Li SM, et al. Cycloplegic refraction by 1%
cyclopentolate in young adults: is it the gold standard? The Anyang
University Students Eye Study (AUSES)[ J]. Br J Ophthalmol, 2018.
doi: 10.1136/bjophthalmol-2018-312199. [Epub ahead of print].
Sanchez I, Ortiz-Toquero S, Martin R . Intrasession repeatability
and intersession reproducibility measurements using VX120
Multidiagnostic unit[ J]. Eye Contact Lens, 2018, 44(Suppl 2):
S266-S272.
电脑自动验光联合综合验光仪在屈光不正中的诊断价值与应用效果. 李奇蒙.智慧健康,2023
Rosenfield M, Chiu NN. Repeatability of subjective and objective
refraction[ J]. Optom Vis Sci, 1995, 72(8): 577-579.