Humphrey视野计的可靠性要求及纳入标准:固视丢失<33%,假阳性和假阴性误差均<25%,视野检测总偏差概率图上至少3个连续点的水平<2%。正常视野的定义:模式标准差(pattern standard deviation,PSD)>5%,且青光眼半侧视野(glaucoma hemifield test,GHT)在正常范围。
Table 3 Repeatability analysis of the measurements of peri-papillary superficial capillaries blood flow parameters at difffferent time points by operator A
视盘上方V D 、下方V D及P D 、颞侧V D及P D的CCC值为0.6270~0.7801,视盘上方P D、鼻侧V D及P D的CCC值为0.5149~0.5847。视盘区V D的C R值为2.0675~4.0630,P D的C R值为0.0447~0.0730;视盘区各区域VD和PD的C V值均<7.2%(3.710%~7.121%,表5)。
表4 不同操作者在同一时间点对正常眼黄斑血流参数测量的再现性分析
Table 4 Reproducibility analysis of the measurements of superficial macular capillaries blood flow parameters at the same time point by difffferent operators
表5 不同操作者在同一时间点对正常眼视盘血流参数测量的再现性分析
Table 5 Reproducibility analysis of the measurements of peri-papillary superficial capillaries blood flow parameters at the same time point by difffferent operators
OCT配有不同的扫描范围、层面分析模式,扫描范围越大,仪器对眼底血流图像的精确分析能力越低[9],因此本研究选择扫描范围3 mm × 3 mm以获取分辨率相对较高和较准确的图像进行分析。当仪器发射出的光波穿透浅层血管到达深层血管并反射回探头时,仪器会把浅层血管的投影(即血管投射伪影)误认为是深层血管,造成血管量化分析误差,因此本研究选择视网膜浅层血管进行检测参数的重复性分析。
Xiao等[10]和赵琦等[11]发现:黄斑3 mm × 3 mm中心、内环的V D、P D测量值的日内重复性良好[ICC=0.87~0.97,CV=(1.2±1.0)%~(4.2±5.5)%],但作者没有研究不同日的重复性及再现性;而在实际临床工作中,眼科医生对患者病情的追踪和随访都不是在同一天,因此本研究选取不同天进行OCTA重复扫描更具有实际意义,也有必要进行间隔时间更长的相关研究。
本研究选取1周内3天的同一时段对黄斑及视盘3 mm × 3 mm血流参数进行重复性和再现性研究,结果显示:非同日的视盘血流参数有较好的重复性及再现性,而黄斑血流参数的重复性及再现性相对较低。与其他作者的前期研究[10-11]相比,本研究黄斑血流参数重复性及再现性略低,可能原因有:1 )眼底图像SS前后不一致。眼底图像质量常以SS来评判,在目前的临床实践和研究中,普遍认为SS≥6代表图像质量好,该图像便被纳入研究中。但多项研究[12-14]发现:SS是影响重复性及再现性的重要因素之一,SS前后差异越大,OCTA参数的可靠性越低。Lei等[14]报道:当SS增加1个单位时,VD增加1.4 mm?1,PD增加0.03。赵琦等[11]报道:黄斑3 mm × 3 mm血流参数重复性好,可能原因是其纳入了SS≥8的图像,而本研究纳入了SS≥7的图像,增大了SS对重复性影响的可能性。但对于视力较差且配合欠佳的患者,采用S S≥8的标准过于苛刻,缺乏实际意义,因此SS≥7相对更符合实际临床工作。Lei等[14]提出在获取OCTA图像时,不仅要提高图像的SS,更要尽量保持前后图像的SS一致,使前后血流参数更具可比性,更有利于在临床中对患者病情的跟踪和随访。2 )OCTA扫描过程中的客观因素多且复杂,而不同日的检查更能代表实际临床工作情况。尽管本研究已尽量控制外界影响因素(如检查时间、室温及环境条件等),但与同一天检查相比,受试者本身的生理变化、情绪波动、配合程度等造成眼底血流的改变较大。同理,非同日条件下操作者的不同生理、心理状态等也会影响测量结果。因此,在OCTA扫描过程中,令受试者闭眼休息、离座走动,必要时提供人工泪液,提高受试者配合程度,可以获取更佳的图像质量,提高重复性及再现性。
正常人眼注视目标时,会产生不易察觉的频繁而有规律的细微运动,称为注视性眼球运动[15]。当以更少的扫描时间和开启FastTracTM图像跟踪功能扫描时,可以减少眼球运动伪影对图像质量造成的影响。本研究所用的Cirrus HD-OCT 5000扫描速度为68 000 A扫/s,理想条件下3 mm × 3 mm范围OCTA扫描时间为5 s,扫描过程中需要患者保持最佳固视状态,但因受试者眨眼、不自觉的眼球抖动造成的血流图像质量欠佳,也有可能导致血流参数的重复性降低。Fenner等[16]也提到运动伪影是影响视网膜浅层血管参数重复性的因素之一,因此,每次扫描前嘱受试者眨眼并对其进行固视宣教是有必要的。目前,OCTA扫描中屈光度的调节是通过操作者目测显示器中眼底图像的清晰度来判断的,其受主观因素影响较大,而眼底图像的清晰度直接影响成像质量[17],这也影响检查结果的重复性及再现性。在扫描过程中,正常受试者的瞳孔在光刺激下收缩,若在扫描过程中因为瞳孔收缩导致眼底入射光线不足在眼底图上形成暗区,对应暗区血流密度下降,因此建议进行OCTA检查前常规散瞳。
目前,对视盘3 mm × 3 mm血流参数重复性及再现性的研究不多。Mastropasqua等[18]报道了视盘4.5 mm × 4.5 mm的VD具有良好的重复性及再现性(ICC=0.767及0.762,CCC=0.6590~0.8891)。本研究发现:视盘3 mm × 3 mm血流参数重复性及再现性优于黄斑(ICC=0.577~0.863 ,CCC=0.5149~0.7801),可能是由于视盘存在视网膜中央动脉及睫状后动脉大血管,其血流较黄斑区稳定,而黄斑区由于中心凹无血管区(fovealavascular zone,FAZ)的存在,整体血流密度明显低于视盘区,血流变化对于黄斑区的影响明显大于视盘区,最终导致视盘区比黄斑区重复性及再现性好。
本研究存在一定的局限性。首先,样本量偏小,由于眼部血流变化与生理及环境等多种因素有关,同时检查结果受操作者及受试者当时心理及生理状态影响,故需要进一步扩大样本量,增加检测指标,进一步排除干扰因素。其次,未评估黄斑FAZ的重复性及再现性,原因是机器自动测量FAZ参数的重复性及再现性不如手动测量,存在一定的系统误差[19],本研究也未分析眼轴和体重指数(body mass index,BMI)对OCTA测量视网膜血流参数的重复性及再现性的影响,将在日后的工作中进一步探讨。
综上所述,非同日的同一时段下黄斑3 mm ×3 mm浅层血流参数有一般或较低的重复性及再现性,可能与受试者生理、心理状态、理解及配合程度、瞳孔大小、运动伪像、图像SS、操作者扫描主观判断有关;视盘3 mm × 3 mm浅层血流参数具有较好的重复性及再现性。提示OCTA的检查结果可供临床医生参考,但仍需结合其他辅助检查综合分析以做出正确的判断。
基金
1、国家自然科学基金面上项目 (81770924,82070963);深圳市科创委基础研究项目 ( JCYJ20190807153005579);
广东省基础与应用基础研究基金项目 (2019A1515011234)。 This work was supported by the National Science and Technology Major Project Foundation
(81770924, 82070963), the Science and Technology Innovation Committee of Shenzhen ( JCYJ20190807153005579) and Guangdong Basic and Applied Basic
Research Foundation (2019A1515011234), China.
参考文献
1、Spaide RF, Klancnik JM Jr, Cooney MJ. Retinal vascular layers imaged
by fluorescein angiography and optical coherence tomography
angiography[ J]. JAMA Ophthalmol, 2015, 133(1): 45-50.
2、朱铁培. 基于光学相干断层扫描血流成像的视网膜微血管网
量化分析及其在糖尿病视网膜病变中的应用[D]. 杭州: 浙江
大学, 2019.
ZHU TP. Quantitative analysis of optical coherence tomography
angiography based retinal microvascular morphology in diabetic
retinopathy[D]. Hangzhou: Zhejiang University, 2019.
3、Harazny JM, Schmieder RE, Welzenbach J, et al. Local application of
tropicamide 0. 5% reduces retinal capillary blood flow[ J]. Blood Press,
2013, 22(6): 371-376.
4、Tang FY, Chan EO, Sun Z, et al. Clinically relevant factors associated
with quantitative optical coherence tomography angiography metrics
in deep capillary plexus in patients with diabetes[ J]. Eye Vis (Lond),
2020, 7: 7.
5、Landis JR, Koch GG. The measurement of observer agreement for
categorical data[ J]. Biometrics, 1977, 33(1): 159-174.
6、Lee JH, Lee MW, Baek SK, et al. Repeatability of manual measurement
of foveal avascular zone area in optical coherence tomography
angiography images in high myopia[ J]. Korean J Ophthalmol, 2020,
34(2): 113-120.
7、Ang M, Tan ACS, Cheung CMG, et al. Optical coherence tomography
angiography: A review of current and future clinical applications[ J].
Graefes Arch Clin Exp Ophthalmol, 2018, 256(2): 237-245.
8、田佳鑫, 王宁利. 原发性开角型青光眼与血流异常的关系[ J]. 中
华实验眼科杂志, 2018, 36(8): 643-648.
TIAN JX, WANG NL. The relationship between primary open
angle glaucoma and blood flow abnormalities[ J]. Chinese Journal of
Experimental Ophthalmology, 2018, 36(8): 643-648.
9、魏文斌, 曾司彦. 重视光相干断层扫描血流成像的临床应用及
其图像的判读[ J]. 中华实验眼科杂志, 2017, 35(10): 865-870.
WEI WB, ZENG SY. Pay ing attention to the clinical
application and image interpretation in optical coherence tomography
angiography[ J]. Chinese Journal of Experimental Ophthalmology,
2017, 35(10): 865-870.
10、Xiao H, Liu X, Liao L, et al. Reproducibility of foveal avascular zone
and superficial macular retinal vasculature measurements in healthy
eyes determined by two different scanning protocols of optical
coherence tomography angiography[ J]. Ophthalmic Res, 2020, 63(3):
244-251.
11、赵琦, 王霄娜, 杨文利, 等. 基于光学微血流成像技术的相干光
断层扫描血流成像对视网膜血流定量分析的可重复性评价[ J].
眼科, 2018, 27(2): 107-110.
ZHAO Q, WANG XN, YANG WL, et al. Repeatability
of quantitative assessment of the retinal microvasculature using
optical coherence tomography angiography based on optical
microangiography[ J]. Ophthalmology in China, 2018, 27(2): 107-110.
12、Lee MW, Nam KY, Lim HB, et al. Long-term repeatability of optical
coherence tomography angiography parameters in healthy eyes[ J].
Acta Ophthalmol, 2020, 98(1): e36-e42.
13、Venugopal JP, Rao HL, Weinreb RN, et al. Repeatability of vessel
density measurements of optical coherence tomography angiography
in normal and glaucoma eyes[ J]. Br J Ophthalmol, 2018, 102(3):
352-357.
14、Lei J, Durbin MK, Shi Y, et al. Repeatability and reproducibility
of superficial macular retinal vessel density measurements using
optical coherence tomography angiography en face images[ J]. JAMA
Ophthalmol, 2017, 135(10): 1092-1098.
15、邸悦, 周行涛, 褚仁远, 等. 注视性眼球运动研究进展[ J]. 中华眼
科杂志, 2012, 48(3): 286-288.
DI Y, ZHOU XT, CHU RY, et al. Research progress of
fixation eye movement[ J]. Chinese Journal of Ophthalmology, 2012,
48(3): 286-288.
16、Fenner BJ, Tan GSW, Tan ACS, et al. Identification of imaging features
that determine quality and repeatability of retinal capillary plexus
density measurements in OCT angiography[ J]. Br J Ophthalmol, 2018,
102(4): 509-514.
17、向金明, 郑琦, 许燕红, 等. Cirrus HD-OCT检测视盘旁视网膜神
经纤维层厚度的可重复性研究[ J]. 中国中医眼科杂志, 2014,
24(4): 262-265.
XIANG JM, ZHENG Q, XU YH, et al. Reproducibility
of thickness measurement of peripapillary retinal nerve fiber layer
with Cirrus HD-OCT in normal eyes[ J]. China Journal of Chinese
Ophthalmology, 2014, 24(4): 262-265.
18、Mastropasqua R, D'Aloisio R, Agnifili L, et al. Functional and structural
reliability of optic nerve head measurements in healthy eyes by means
of optical coherence tomography angiography[ J]. Medicina (Kaunas),
2020, 56(1): 44.
19、Lin A, Fang D, Li C, et al. Reliability of foveal avascular zone metrics
automatically measured by Cirrus optical coherence tomography
angiography in healthy subjects[ J]. Int Ophthalmol, 2020, 40(3):
763-773.